Read Selecting Best Probe in 5 Minutes App Note
"img-power-probes-with-mda"
">
Read Selecting Best Probe in 5 Minutes App Note
">

性能最佳的高压探头

Read Selecting Best Probe in 5 Minutes App Note
">

Teledyne LeCroy 高压探头以其卓越的共模抑制比 (CMRR) 和信号保真度以及高精度而著称。 电流探头和电流传感器适配器也可用于提供精确的交流、直流和脉冲电流测量。

阅读在 5 分钟内选择最佳探头的应用说明
">img-power-probes-with-mda 的
Explore more
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series (Si) or DL-ISO Series (SiC/GaN)
Might have better performance but maybe not worth high price. CMRR advantage not as helpful at 48 Vdc

Conduction loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.

Switching Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Passive Probe supplied with Oscilloscope
Suitable for use on low-side device only
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-HCM Series
Good performance, best value
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series (Si) or DL-ISO Series (SiC/GaN)
Cost is very high for capability provided
Probe noise will be higher than DL-HCM at this voltage range

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Series/Shunt Resistor

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Cannot measure voltage drop across resistor with ground-referenced probe.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
AP033 or DL-HCM Series
Probes with gain (AP033) may perform acceptably but may not meet common-mode needs. Otherwise, performance likely borderline ok.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series (Si) or DL-ISO Series (SiC/GaN)
Low signal amplitude may still result in marginal measurement.

Sensor Signal

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-HCM Series
Low attenuation means low additive noise with very small signals.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series (Si) or DL-ISO Series (SiC/GaN)
Benefits of optical isolation less apparent and less required at lower bus voltages.

Discrete Components

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-HCM Series
Best solution in all cases
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series (Si) or DL-ISO Series (SiC/GaN)
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Line Side (AC) Input

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope
Check ratings of probe for suitability. Line-neutral (ground-referenced) voltage probing only – cannot probe line-line.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC Bus/Link

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope
High attenuation, could be noisy
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-HCM Series
Low attenuation means low additive noise with very small signals.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

Inverter/Drive PWM Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope
Check ratings of probe for suitability. Line-reference voltage probing only – cannot probe line-line. Board REF must be able to be connected to scope GND.
Best option
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
DL-HCM Series
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC-DC Converter HV Input/Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope
Check ratings of probe for suitability. High attenuation, could be noisy. Board REF must be able to be connected to scope GND.
Best option
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
DL-HCM Series
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Gate drive

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Passive Probe Supplied with Oscilloscope, PPE or HVP Series
Low-side gate drive signals only. Board REF = Scope GND. High attenuation = more noise.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD310x models)
Circuit loading may be an issue in some cases.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO Series
Lowest tip capacitance and loading, best CMRR.

Conduction loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.

Switching Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
PPE or HVP Series
Suitable for use on low-side device only.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Good performance, best value.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
DL-ISO Series
Cost is very high for capability provided.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Series/Shunt Resistor

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Cannot measure voltage drop across resistor with ground-referenced probe.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
HVD Series (HVD310x models)
Probe attenuation too high and signal amplitude likely too low for a good measurement.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Low signal amplitude may still result in marginal measurement.

Sensor Signal

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD310x models)
High attenuation means high additive noise with very small signals.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO Series
Lower attenuation will generally result in lower additive noise.

Discrete Components

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Best solution in all cases
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Line Side (AC) Input

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-neutral (ground-referenced) voltage probing only – cannot probe line-line.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC Bus/Link

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

Inverter/Drive PWM Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-reference voltage probing only – cannot probe line-line. Board REF must be able to be connected to scope GND.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC-DC Converter HV Input/Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy. Board REF must be able to be connected to scope GND.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive HV Single-ended,
Ground-referenced

Active HV Differential,
High-attenuation isolated

Active
Single-ended,
Fiber-optic Isolated probe

Gate Drive

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Passive Probe Supplied with Oscilloscope, PPE or HVP Series
Low-side gate drive signals only. Board REF = Scope GND. High attenuation = more noise.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
Circuit loading may be an issue in some cases. Bandwidth likely too low for GaN.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-ISO Series
Lowest tip capacitance and loading, best CMRR.

Conduction Loss

do not use
Single-ended ground-referenced design will limit utility in this application.
do not use
Single-ended ground-referenced design will limit utility in this application.
do not use
Single-ended ground-referenced design will limit utility in this application.

Switching Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
PPE or HVP Series
Suitable for use on low-side device only.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
Circuit loading may be an issue in some cases. Bandwidth likely too low for GaN.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-ISO Series
Best probe CMRR at high frequencies is beneficial for fast WBG slew rates

 
 
Application

Passive HV Single-ended,
Ground-referenced

Active HV Differential,
High-attenuation isolated

Active
Single-ended,
Fiber-optic Isolated probe

Series/Shunt Resistor

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Cannot measure voltage drop across resistor with ground-referenced probe.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
HVD Series (HVD3220)
Probe attenuation too high and signal amplitude likely too low for a good measurement.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Low signal amplitude may still result in marginal measurement.

Sensor Signal

do not use
Single-ended ground-referenced design will limit utility in this application.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
High attenuation means high additive noise with very small signals.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO Series
Lower attenuation will generally result in lower additive noise.

Discrete Components

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3220)
Best solution in all cases
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Line Side (AC) Input

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-neutral (ground-referenced) voltage probing only – cannot probe line-line.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x or HVD32xx)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC Bus/Link

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability
High attenuation, could be noisy
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x or HVD32xx)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

Inverter/Drive PWM Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-reference voltage probing only – cannot probe line-line. Board REF must be able to be connected to scope GND.
Best option
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
DL-HCM Series
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC-DC Converter HV Input/Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy. Board REF must be able to be connected to scope GND.
Best option
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
HVD Series (HVD310x or HVD32xx)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Gate drive

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Passive Probe Supplied with Oscilloscope, PPE or HVP Series
Low-side gate drive signals only. Board REF = Scope GND. High attenuation = more noise.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD310x models)
Circuit loading may be an issue in some cases.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO Series
Lowest tip capacitance and loading, best CMRR.

Conduction loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.

Switching Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
PPE or HVP Series
Suitable for use on low-side device only.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD310x models)
Good performance, best value.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
DL-ISO Series
Cost is very high for capability provided.

 
 
Application

Passive HV Single-ended,
Ground-referenced

Active HV Differential,
High-attenuation isolated

Active
Single-ended,
Fiber-optic Isolated probe

Series/Shunt Resistor

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Cannot measure voltage drop across resistor with ground-referenced probe.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
HVD Series (HVD3206 or HVD3220)
Probe attenuation too high and signal amplitude likely too low for a good measurement.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Low signal amplitude may still result in marginal measurement.

Sensor Signal

do not use
Single-ended ground-referenced design will limit utility in this application.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
High attenuation means high additive noise with very small signals.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO Series
Lower attenuation will generally result in lower additive noise.

Discrete Components

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3220)
Best solution in all cases
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Line Side (AC) Input

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-neutral (ground-referenced) voltage probing only – cannot probe line-line.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC Bus/Link

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability
High attenuation, could be noisy
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3206 or HVD3220)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

Inverter/Drive PWM Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-reference voltage probing only – cannot probe line-line. Board REF must be able to be connected to scope GND.
Best option
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
HVD Series (HVD3206 or HVD3220)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC-DC Converter HV Input/Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy. Board REF must be able to be connected to scope GND.
Best option
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
HVD Series (HVD3206 or HVD3220)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active HV Differential,
High-attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Gate drive

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Passive Probe Supplied with Oscilloscope, PPE or HVP Series
Low-side gate drive signals only. Board REF = Scope GND. High attenuation = more noise.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
Circuit loading may be an issue in some cases. Bandwidth likely too low for GaN.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-ISO Series
Lowest tip capacitance and loading, best CMRR.

Conduction loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.

Switching Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
PPE or HVP Series
Suitable for use on low-side device only.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
Circuit loading may be an issue in some cases
Bandwidth likely too low for GaN
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
DL-ISO Series
Best probe CMRR at high frequencies is beneficial for fast WBG slew rates

 
 
Application

Passive HV Single-ended,
Ground-referenced

Active HV Differential,
High-attenuation isolated

Active
Single-ended,
Fiber-optic Isolated probe

Series/Shunt Resistor

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Cannot measure voltage drop across resistor with ground-referenced probe.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
HVD Series (HVD3220)
Probe attenuation too high and signal amplitude likely too low for a good measurement.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Low signal amplitude may still result in marginal measurement.

Sensor Signal

do not use
Single-ended ground-referenced design will limit utility in this application.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3220)
High attenuation means high additive noise with very small signals.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO Series
Lower attenuation will generally result in lower additive noise.

Discrete Components

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3220)
Best solution in all cases
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO Series or DL-ISO Series
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Line Side (AC) Input

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-neutral (ground-referenced) voltage probing only – cannot probe line-line.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC Bus/Link

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3206 or HVD3220)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

Inverter/Drive PWM Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. Line-reference voltage probing only – cannot probe line-line. Board REF must be able to be connected to scope GND.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3206 or HVD3220)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

DC-DC Converter HV Input/Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
Passive Probe Supplied with Oscilloscope, PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy. Board REF must be able to be connected to scope GND.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3206 or HVD3220)
Best solution in all cases
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Gate Drive

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Passive Probe Supplied with Oscilloscope, PPE or HVP Series
Low-side gate drive signals only. Board REF = Scope GND. High attenuation = more noise.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3605)
Circuit loading may be an issue in some cases. Attenuation will be high, resulting in more noise in the measurement. Bandwidth might be too low for SiC.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO or DL-ISO Series
Lowest tip capacitance and loading, best CMRR.

Conduction Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.

Switching Loss

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
PPE or HVP Series
Check ratings of probe for suitability. Suitable for use on low-side device only.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3605)
Good performance, best value. Be sure Vpk-pk requirement doesn’t exceed rated range.
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Voltage range required generally exceeds probe capabilities.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
Differential,
Attenuation Isolated probe

Active
Single-ended,
Fiber-optic Isolated probe

Series/Shunt Resistor

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Cannot measure voltage drop across resistor with ground-referenced probe.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
HVD Series (HVD3605)
Probe attenuation too high and signal amplitude likely too low for a good measurement.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO or DL-ISO Series
Low signal amplitude may still result in marginal measurement.

Sensor Signal

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
safe but not suitable
While the probe will provide a result and will not be damaged inmaking the measurement, most users would find the probe does not work well in this application.
HVD Series (HVD3605)
Very high attenuation means very high additive noise with very small signals.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVFO or DL-ISO Series
Lower probe attenuation will generally result in lower additive noise.

Discrete Components

do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Single-ended ground-referenced design will limit utility in this application.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3605)
Best solution in all cases
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVFO or DL-ISO Series
Cost is very high for capability provided. Voltage range required may exceed probe capability.

 
 
Application

Passive
HV Single-ended,
Ground referenced probe

Active
HV Differential,
High-attenuation isolated

Active
Single-ended,
Fiber-optic Isolated probe

Line Side (AC) Input

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
PPE Series, or HVP Series
Check ratings of probe for suitability. Line-neutral (ground-referenced) voltage probing only – cannot probe line-line.
caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
HVD Series (HVD3605)
Be sure that Vpk-pk requirement doesn’t exceed differential voltage range
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Voltage range required generally exceeds probe capabilities.

DC Bus/Link

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3605)
Best solution in all cases
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Voltage range required generally exceeds probe capabilities.

Inverter/Drive PWM Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
PPE Series, or HVP Series
Check ratings of probe for suitability. Line-reference voltage probing only – cannot probe line-line. Board REF must be able to be connected to scope GND.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3605)
Best solution in all cases
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Voltage range required generally exceeds probe capabilities.

DC-DC Converter HV Input/Output

caution
There are some compromises in the performance of the probe in this application, though some users may find the probe works fine for them.
PPE Series, or HVP Series
Check ratings of probe for suitability. High attenuation, could be noisy. Board REF must be able to be connected to scope GND.
Best option
This is the perfect probe for the application. There are few issues with its use, and it has been optimized in price and performance for this application.
HVD Series (HVD3605)
Best solution in all cases
do not use
The probe should absolutely not be used in this application as damage to the probe, oscilloscope or device under test (DUT) may occur, or harm may come to the operator.
Voltage range required generally exceeds probe capabilities.
">

高压探头选择指南 - 在下面进行选择

重设

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

栅极驱动

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器提供的无源探头
仅低侧栅极驱动信号。 电路板 REF = ScopeGND。 高衰减=更多噪音。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-HCM系列
具有非常合理的 CMRR 和低尖端电容的可能的最佳整体性能
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列 (Si) 或 DL-ISO 系列 (SiC/GaN)
可能有更好的性能,但可能不值高价。 CMRR 优势在 48 Vdc 时没有帮助

传导损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。

开关损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器提供的无源探头
仅适用于低侧设备
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-HCM系列
良好的性能,最佳价值
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列 (Si) 或 DL-ISO 系列 (SiC/GaN)
所提供能力的成本非常高
在此电压范围内,探头噪声将高于 DL-HCM

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

串联/分流电阻

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
无法使用接地参考探头测量电阻两端的压降。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
AP033 或 DL-HCM 系列
具有增益的探头 (AP033) 的性能可能可以接受,但可能无法满足共模需求。 否则,性能可能接近正常。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列 (Si) 或 DL-ISO 系列 (SiC/GaN)
低信号幅度仍可能导致边际测量。

传感器信号

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-HCM系列
低衰减意味着非常小的信号具有低附加噪声。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列 (Si) 或 DL-ISO 系列 (SiC/GaN)
在较低的总线电压下,光隔离的好处不太明显,也不太需要。

分立元件

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-HCM系列
所有情况下的最佳解决方案
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列 (Si) 或 DL-ISO 系列 (SiC/GaN)
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

线路侧 (AC) 输入

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器提供的无源探头
检查探头的额定值是否适用。 仅中性线(接地参考)电压探测 – 不能探测线路。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD310x)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

直流母线/链接

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器提供的无源探头
高衰减,可能会有噪音
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-HCM系列
低衰减意味着非常小的信号具有低附加噪声。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

逆变器/驱动器 PWM 输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器提供的无源探头
检查探头的额定值是否适用。 仅线路参考电压探测——不能探测线路。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
DL-HCM系列
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

DC-DC转换器高压输入/输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
DL-HCM系列
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

栅极驱动

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器、PPE 或 HVP 系列提供的无源探头
仅低侧栅极驱动信号。 电路板 REF = 范围 GND。 高衰减=更多噪音。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列(HVD310x 型号)
在某些情况下,电路负载可能是一个问题。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压振荡器系列
最低尖端电容和负载,最佳 CMRR。

传导损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。

开关损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
PPE 或 HVP 系列
仅适用于低侧设备。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
良好的性能,最佳价值。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
DL-ISO系列
所提供功能的成本非常高。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

串联/分流电阻

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
无法使用接地参考探头测量电阻两端的压降。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
HVD 系列(HVD310x 型号)
探头衰减太高,信号幅度可能太低,无法进行良好的测量。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
低信号幅度仍可能导致边际测量。

传感器信号

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列(HVD310x 型号)
高衰减意味着带有非常小信号的高加性噪声。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压振荡器系列
较低的衰减通常会导致较低的加性噪声​​。

分立元件

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
所有情况下的最佳解决方案
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
高压振荡器系列
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

线路侧 (AC) 输入

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅中性线(接地参考)电压探测 – 不能探测线路。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

直流母线/链接

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

逆变器/驱动器 PWM 输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅线路参考电压探测——不能探测线路。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

DC-DC转换器高压输入/输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

无源高压单端,
以地为参考

有源高压差速器,
高衰减隔离

活跃
单端,
光纤隔离探头

栅极驱动

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器、PPE 或 HVP 系列提供的无源探头
仅低侧栅极驱动信号。 电路板 REF = 范围 GND。 高衰减=更多噪音。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
在某些情况下,电路负载可能是一个问题。 带宽对于 GaN 来说可能太低了。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-ISO系列
最低尖端电容和负载,最佳 CMRR。

传导损耗

不使用
单端接地参考设计将限制此应用中的效用。
不使用
单端接地参考设计将限制此应用中的效用。
不使用
单端接地参考设计将限制此应用中的效用。

开关损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
PPE 或 HVP 系列
仅适用于低侧设备。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
在某些情况下,电路负载可能是一个问题。 带宽对于 GaN 来说可能太低了。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-ISO系列
高频下的最佳探头 CMRR 有利于快速 WBG 转换速率

 
 
应用

无源高压单端,
以地为参考

有源高压差速器,
高衰减隔离

活跃
单端,
光纤隔离探头

串联/分流电阻

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
无法使用接地参考探头测量电阻两端的压降。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
HVD 系列 (HVD3220)
探头衰减太高,信号幅度可能太低,无法进行良好的测量。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
低信号幅度仍可能导致边际测量。

传感器信号

不使用
单端接地参考设计将限制此应用中的效用。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
高衰减意味着带有非常小信号的高加性噪声。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压振荡器系列
较低的衰减通常会导致较低的加性噪声​​。

分立元件

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3220)
所有情况下的最佳解决方案
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

线路侧 (AC) 输入

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅中性线(接地参考)电压探测 – 不能探测线路。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 或 HVD32xx)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

直流母线/链接

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用
高衰减,可能会有噪音
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 或 HVD32xx)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

逆变器/驱动器 PWM 输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅线路参考电压探测——不能探测线路。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
DL-HCM系列
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

DC-DC转换器高压输入/输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
HVD 系列(HVD310x 或 HVD32xx)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

栅极驱动

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器、PPE 或 HVP 系列提供的无源探头
仅低侧栅极驱动信号。 电路板 REF = 范围 GND。 高衰减=更多噪音。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列(HVD310x 型号)
在某些情况下,电路负载可能是一个问题。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压振荡器系列
最低尖端电容和负载,最佳 CMRR。

传导损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。

开关损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
PPE 或 HVP 系列
仅适用于低侧设备。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD310x 型号)
良好的性能,最佳价值。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
DL-ISO系列
所提供功能的成本非常高。

 
 
应用

无源高压单端,
以地为参考

有源高压差速器,
高衰减隔离

活跃
单端,
光纤隔离探头

串联/分流电阻

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
无法使用接地参考探头测量电阻两端的压降。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
HVD 系列(HVD3206 或 HVD3220)
探头衰减太高,信号幅度可能太低,无法进行良好的测量。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
低信号幅度仍可能导致边际测量。

传感器信号

不使用
单端接地参考设计将限制此应用中的效用。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
高衰减意味着带有非常小信号的高加性噪声。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压振荡器系列
较低的衰减通常会导致较低的加性噪声​​。

分立元件

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3220)
所有情况下的最佳解决方案
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

线路侧 (AC) 输入

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅中性线(接地参考)电压探测 – 不能探测线路。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压系列
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

直流母线/链接

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用
高衰减,可能会有噪音
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD3206 或 HVD3220)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

逆变器/驱动器 PWM 输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅线路参考电压探测——不能探测线路。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
HVD 系列(HVD3206 或 HVD3220)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

DC-DC转换器高压输入/输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
HVD 系列(HVD3206 或 HVD3220)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

有源高压差速器,
高衰减隔离探头

活跃
单端,
光纤隔离探头

栅极驱动

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器、PPE 或 HVP 系列提供的无源探头
仅低侧栅极驱动信号。 电路板 REF = 范围 GND。 高衰减=更多噪音。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
在某些情况下,电路负载可能是一个问题。 带宽对于 GaN 来说可能太低了。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-ISO系列
最低尖端电容和负载,最佳 CMRR。

传导损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。

开关损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
PPE 或 HVP 系列
仅适用于低侧设备。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
在某些情况下,电路负载可能是一个问题
GaN 的带宽可能太低
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
DL-ISO系列
高频下的最佳探头 CMRR 有利于快速 WBG 转换速率

 
 
应用

无源高压单端,
以地为参考

有源高压差速器,
高衰减隔离

活跃
单端,
光纤隔离探头

串联/分流电阻

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
无法使用接地参考探头测量电阻两端的压降。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
HVD 系列 (HVD3220)
探头衰减太高,信号幅度可能太低,无法进行良好的测量。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
低信号幅度仍可能导致边际测量。

传感器信号

不使用
单端接地参考设计将限制此应用中的效用。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3220)
高衰减意味着带有非常小信号的高加性噪声。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压振荡器系列
较低的衰减通常会导致较低的加性噪声​​。

分立元件

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3220)
所有情况下的最佳解决方案
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 系列或 DL-ISO 系列
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

线路侧 (AC) 输入

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅中性线(接地参考)电压探测 – 不能探测线路。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
高压系列
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

直流母线/链接

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD3206 或 HVD3220)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

逆变器/驱动器 PWM 输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 仅线路参考电压探测——不能探测线路。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD3206 或 HVD3220)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

DC-DC转换器高压输入/输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
随示波器、PPE 系列或 HVP 系列提供的无源探头
检查探头的额定值是否适用。 高衰减,可能会有噪音。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列(HVD3206 或 HVD3220)
所有情况下的最佳解决方案
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

栅极驱动

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
随示波器、PPE 或 HVP 系列提供的无源探头
仅低侧栅极驱动信号。 电路板 REF = 范围 GND。 高衰减=更多噪音。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3605)
在某些情况下,电路负载可能是一个问题。 衰减会很高,从而导致测量中出现更多噪声。 对于 SiC,带宽可能太低。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVFO 或 DL-ISO 系列
最低尖端电容和负载,最佳 CMRR。

传导损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。

开关损耗

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
PPE 或 HVP 系列
检查探头的额定值是否适用。 仅适用于低侧设备。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3605)
良好的性能,最佳价值。 确保 Vpk-pk 要求不超过额定范围。
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
所需的电压范围通常超过探头的能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
微分,
衰减隔离探头

活跃
单端,
光纤隔离探头

串联/分流电阻

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
无法使用接地参考探头测量电阻两端的压降。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
HVD 系列 (HVD3605)
探头衰减太高,信号幅度可能太低,无法进行良好的测量。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 或 DL-ISO 系列
低信号幅度仍可能导致边际测量。

传感器信号

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
安全但不适合
虽然探头会提供结果并且在进行测量时不会损坏,但大多数用户会发现探头在这种应用中不能很好地工作。
HVD 系列 (HVD3605)
非常高的衰减意味着非常高的加性噪声​​和非常小的信号。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVFO 或 DL-ISO 系列
较低的探头衰减通常会导致较低的加性噪声​​。

分立元件

不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
单端接地参考设计将限制此应用中的效用。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3605)
所有情况下的最佳解决方案
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVFO 或 DL-ISO 系列
所提供功能的成本非常高。 所需的电压范围可能超过探头能力。

 
 
应用

被动
高压单端,
接地参考探头

活跃
高压差速器,
高衰减隔离

活跃
单端,
光纤隔离探头

线路侧 (AC) 输入

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
PPE 系列或 HVP 系列
检查探头的额定值是否适用。 仅中性线(接地参考)电压探测 – 不能探测线路。
警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
HVD 系列 (HVD3605)
确保 Vpk-pk 要求不超过差分电压范围
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
所需的电压范围通常超过探头的能力。

直流母线/链接

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
PPE 系列或 HVP 系列
检查探头的额定值是否适用。 高衰减,可能会有噪音
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3605)
所有情况下的最佳解决方案
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
所需的电压范围通常超过探头的能力。

逆变器/驱动器 PWM 输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
PPE 系列或 HVP 系列
检查探头的额定值是否适用。 仅线路参考电压探测——不能探测线路。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3605)
所有情况下的最佳解决方案
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
所需的电压范围通常超过探头的能力。

DC-DC转换器高压输入/输出

警告
在此应用程序中,探针的性能存在一些折衷,但一些用户可能会发现该探针对他们来说工作正常。
PPE 系列或 HVP 系列
检查探头的额定值是否适用。 高衰减,可能会有噪音。 电路板 REF 必须能够连接到示波器 GND。
最佳选择
这是适合该应用的完美探头。 它的使用几乎没有问题,并且针对此应用程序在价格和性能方面进行了优化。
HVD 系列 (HVD3605)
所有情况下的最佳解决方案
不使用
绝对不应在此应用中使用探头,因为可能会损坏探头、示波器或被测设备 (DUT),或者可能对操作员造成伤害。
所需的电压范围通常超过探头的能力。
Choose Category
">
Explore
 

Superior Performance

  • 1 kV, 2 kV, 6 kV CAT safety rated models
  • Widest differential voltage ranges available – up to 7.6 kV
  • ProBus interface and wide oscilloscope compatibility

Exceptional CMRR and Bandwidth

  • 65 dB CMRR at 1 MHz - 50x better than competitive probes
  • Up to 400 MHz bandwidth
  • Best signal fidelity - lowest noise and overshoot

Highest Precision

  • Best gain accuracy – 1%
  • Autozero with auto disconnect switch
  • High offset capability at both high and low attenuation
">
Explore
">

高压差分探头 – HVD 系列

HVD3000A 系列高压差分探头在宽频率范围内提供高 CMRR,以简化在嘈杂、高共模电力电子环境中发现的测量挑战。 该探头的设计易于使用,可实现安全、精确的高压浮地测量。

Explore
 

出众的性能

  • 1 kV、2 kV、6 kV CAT 安全额定型号
  • 可用的最宽差分电压范围 – 高达 7.6 kV
  • ProBus 接口和广泛的示波器兼容性

出色的共模抑制比和带宽

  • 1 MHz 时实现 65 dB 的共模抑制比 - 比竞争对手的探头好 50 倍
  • 高达 400 MHz 的带宽
  • 最佳的信号保真度 - 最低的噪声和过冲

更高的精度

  • 最佳增益精度 – 1%
  • 带自动断路开关的自动归零
  • 在高衰减比和低衰减比下都具备的高偏置能力
Explore
 

Highest accuracy

  • 1.5% system accuracy
  • Precision Gain Calibration
  • High 160 dB CMRR

Best Signal Fidelity

  • Most faithful representation of the signal
  • Low probe loading and minimal overshoot
  • 435 ps rise time

Probe Compare: DL-ISO vs. Tek IsoVu Video

Probe Compare Setup Video

Comprehensive Connectivity

  • MMCX for high bandwidth convenience
  • Square headers for 2.5 kV performance
  • Very flexible tips vs Tektronix IsoVu

View DL-ISO Probe Tips – Most Flexible Video

">
Explore
">

高压光隔离示波器探头 – DL-ISO 系列

DL-ISO 1GHz 探头以最高的精度、最佳的信号保真度和全面的连接性,为 GaN 和 SiC 器件表征提供最高的置信度。

Explore
 

更高的精度

  • 1.5% 系统精度
  • 精确的增益校准
  • 160 dB 高共模抑制比
Probe Compare: DL-ISO vs. Tek IsoVu Video

Probe Compare Setup Video

" data-gt-human-content="true">

更好的信号保真度

  • 信号的最真实呈现
  • 低探头负载和最小的过冲
  • 435 ps 上升时间

View DL-ISO Probe Tips – Most Flexible Video

">

全面的连接

View DL-ISO Probe Tips – Most Flexible Video

" data-gt-human-content="true">
  • MMCX 提供高带宽便利性
  • 用于 2.5 kV 性能的方形接头
  • 与泰克 IsoVu 相比更加灵活的前端

查看 DL-ISO 探头前端视频 – 最灵活的方式

Explore
 

Optical Isolation for Silicon MOSFETs

  • 150 MHz bandwidth
  • 35 kV common-mode voltage rating
  • Optical isolation for reduced loading of DUT

Superior Noise Rejection

  • 140 dB CMRR
  • Low loop inductance
  • Low attenuation

Simple, Compact, Affordable

  • Fits into tight spaces
  • Fits into tight budgets
  • No compromise on performance
">
Explore
">

高压光隔离示波器探头 – HVFO 系列

具有 108 MHz 带宽的新型 HVFO150 探头, 其适用于硅和碳化硅设计。 专为测量漂浮在高电压总线上的小信号而设计,提供卓越的性能与高共模抑制比,并运用光隔离技术以减少 DUT 的负载。

Explore
 

为硅 MOSFET 提供光隔离

  • 150 MHz 带宽
  • 35 kV 共模电压
  • 用于减少 DUT 负载的光隔离技术

卓越的噪声抑制

  • 140 dB 共模抑制比
  • 低回路电感
  • 低衰减比

简单、紧凑、实惠

  • 适用于狭小的探测空间
  • 适合于紧张的预算
  • 性能不妥协
Explore
"dl-hcm-probe-image"

Ideal for 48 V GaN

  • 250 MHz, 500 MHz, and 1 GHz bandwidth
  • 80 V dynamic range
  • 60 V common mode

Highest accuracy, Lowest Noise, Highest Rejection

  • 0.5% gain accuracy
  • Precision Gain Calibration
  • Best LF flatness (0.1 dB)
  • Exceptional CMRR out to high frequencies

Wide variety of tips

  • Wide variety of tips standard
  • ProBus active probe interface
  • Optional high temperature solder-in tip available
">
Explore
">

60 V 共模差分探头 – DL-HCM 系列

60 V 共模差分探头是低压 GaN 功率转换测量的理想探头,具有最高的精度、最佳的 CMRR 和最低的噪声。 它们也是理想的汽车电源探头,用于测试 LV148、LV124、LV123、ISO 16750 和 ISO 7637 汽车标准,以测量电池电压。

Explore
dl-hcm-探针图像

48 V GaN 的理想选择

  • 250 MHz、500 MHz 和 1 GHz 带宽
  • 80 V 动态范围
  • 60 V 共模

最高的精度、最低的噪声、最高的共模抑制

  • 0.5% 增益精度
  • 精密增益校准
  • 最佳低频平坦度 (0.1 dB)
  • 卓越的 CMRR 输出至高频

超多类型的尖端

  • 多种尖端标准
  • ProBus 有源探头接口
  • 提供可选的高温焊入式尖端
Explore
"dl-hcm-probe-image"

Ideal for Ground-referenced Signals

  • 500 MHz bandwidth
  • 1000 VRMS input (CAT II)
  • 6000 VPK transients (CAT I)

Standard Accessories Include:

  • Sprung hook
  • BNC adapter
  • Insulating tip
  • Swivel ground lead and lead adapter

HV Passive Probe Applications

  • Surge/EFT testing
  • Lower-side gate-drive signals
  • General ground-referenced HV use
">
Explore
">

高压无源探头

Teledyne LeCroy HV 无源探头可安全探测 6000 VPK瞬态并可承受高达 1000 V 的电压 有效值(1500V DC)。 它们提供高频响应以实现更快的上升时间,非常适合探测以地为参考的高压信号。

Explore
dl-hcm-探针图像

非常适合接地参考信号

  • 500 MHz 带宽
  • 1000 V 有效值输入(CAT II)
  • 6000 V PK瞬态(CAT I)

标准配件包括:

  • 弹簧钩
  • BNC适配器
  • 绝缘头
  • 旋转接地线和导线适配器

高压无源探头应用

  • 浪涌/EFT 测试
  • 下侧栅极驱动信号
  • 一般接地参考高压使用
Explore
 

A current probe for every need

  • 30 to 500 A continuous current
  • Up to 100 MHz bandwidth
  • Powered from the oscilloscope, no external power supply required

Best performance

  • High sensitivity of 1 mA/div for low-current measurements
  • No breaking of circuit or insertion of shunt to make measurement
  • Autozero and degauss capabilities built into oscilloscope

Using a current sensor?

  • CA10 enables 3rd party probe to operate like Teledyne LeCroy probe
  • Programmable EEPROM saves 3rd party current sensor parameters
  • Easy to use, saves time and possible errors

Learn more

">
Explore
">

电流探头

Teledyne LeCroy 电流探头不需要断开电路或插入分流器即可进行准确可靠的电流测量。 基于霍尔效应和变压器技术的组合,Teledyne LeCroy 电流探头非常适合进行精确的交流、直流和脉冲电流测量。

Explore
 

满足各种需求的电流探头

  • 30 至 500 A 连续电流
  • 高达 100 MHz 的带宽
  • 由示波器供电,无需外部电源

更好的性能

  • 1 mA/div 的高灵敏度,适用于小电流测量
  • 无需断开电路或插入分流器即可进行测量
  • 示波器内置自动归零和消磁功能
Learn more

">

使用电流传感器?

Learn more

" data-gt-human-content="true">
  • CA10 使第三方探头能够像 Teledyne LeCroy 探头一样运行
  • 可编程 EEPROM 可保存第三方电流传感器参数
  • 易于使用,节省时间和可能的错误

了解更多

Sign Up for Updates
">
Sign Up for Updates, be Contacted by Sales, or would like to Request a Demo
">

需要帮助或信息?

请填写此表格,让我们知道您是否愿意 订阅活动快讯,同 销售人员联系, 或者 索取方案演示